Event or Problem Description
|
|
While at customer site for a preventive maintenance, a field service engineer reported a ¿smoke¿ or mist/haze emitting from the sterrad® nx sterilizer.There was no report of any injuries or human reactions.This event is being reported as a malfunction report subsequent to a previous serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service engineer replaced the oil mist filter, catalytic converter, solenoid valve, and vacuum pump to resolve the mist/haze issue.Unit meets specifications and was returned to service.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by advanced sterilization products, or its employees that the report constitutes an admission that the product, advanced sterilization products, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Asp complaint ref #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the mist/haze issue, system risk analysis (sra), and photo analysis.¿ trending analysis of the mist/haze issue for the sterrad® nx unit was reviewed for the prior six months from the event date, and no significant trend was observed.¿ review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for further analysis.Field service provided a photo of the oil mist filter for review, and the filter was noted to exhibit burning/charring along the outside of the body of the filter.The failure was confirmed by photo analysis.The assignable cause of the mist/haze is likely due to the oil mist filter, catalytic converter, solenoid valve, and vacuum pump.The asp field service engineer replaced the parts and confirmed the sterrad® nx met functional specifications after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
|
| |
|