|
An urgent medical device recall (umdr) notice was sent to the us customers on 21-jun-2023.The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.Customers are instructed to contact ascensia diabetes care customer service to return the affected product and receive the replacement product.The umdr explains the potential for incorrect units of measurement associated with the contour next gen meters.(b)(4) has implemented improvements to resolve the issue of incorrect units of measurement.The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in sections a2 and a4 as the customer's age and weight were not provided.Since the issue occurred at (b)(4) which is a packaging site for products distributed in the us, (b)(4) address has been captured in section g1 (continued).Since the issue was associated with the packaging of the device, the device manufacture date in section h4 has been captured as 08-feb-2023 when the suspected device was packaged at (b)(4).Recall was checked off in section h7, as this issue is associated with the umdr and correction/removal number has been updated in section h9.
|