| Catalog Number |
010000849 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the procedure, the liner would not assemble with the shell.The procedure was completed with another device.There was no harm or health consequences to the patient.Attempts have been made and no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.One g7 neutral e1 liner 32mm e item# 010000849 lot# 6742238 was returned and evaluated.Upon visual inspection the locking feature of the device is deformed and there are circular indentations to the top of the od.Event confirmed based on the evaluation of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.A definitely root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|