| Catalog Number |
42800-27 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a transpac¿ it monitoring kit, 60" safeset reservoir, 2 blood sampling port, 3ml flush device, macrodrip squeeze flush mechanism appears ¿turned¿ and when the registered nurse (rn) tried to straighten it, it began leaking from the flush mechanism area.The product wa used for one hour.There was patient involved but no harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation; however, testing has not yet been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint of leakage was not confirmed on the returned set.No visual anomalies were observed on the returned set.When the returned set was primed and pressure leak tested no leaks were observed near the flush mechanism.The set had performed per the specifications.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.Additional information in g1.
|
| |
|