| Model Number |
UM-G20-29R |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that the tip of the ultrasonic probe was caught in the guide wire and fell off into the patient's duodenum during a diagnostic intraductal ultrasonography.The procedure was completed with the device, and at the physician's discretion, the tip was not collected.The fallen piece can be expected to be expelled from the patient's body.There were no reports of patient injury or any adverse events.There is no evidence that a life-threatening event occurred, that the patient developed permanent damage or impairment, or that additional medical/surgical intervention was done to prevent permanent damage or impairment.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The g2 field was corrected based on information available at the time of the initial submission.The following information was provided by the facility: the doctor at the facility confirmed by endoscopy that foreign body that had dropped out of the body.The facility confirmed that the foreign body was not confirmed to be the one of the ultrasonic probes of subject device.The foreign body may be a product of another company.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's complaint was not confirmed.Based on the results of the investigation, there was no breakage of the tip pf the ultrasonic probe.In addition, the following non-reportable malfunctions were found during the device evaluation: indentations and slight twists on the distal sheath, and the guide wire insertion port was chipped and indented.Also, the ultrasound image was not visualized properly.The malfunctions were due to physical stress applied by the user.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|