INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; N/A Back to Search Results
Catalog Number A1114
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported that the torque screw on mayfield infinity skull clamp (a1114) does not exert pressure, and the force indicator of the adjustment screw does not allow to visualize the value of the force applied in increments of 20 lbs.The failure occurred in surgery with a patient, which caused an adverse event.However, no surgical delay nor patient injury reported.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The mayfield infinity skull clamp (a1114) was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined; however, based on the reported complaint, probable root cause is damage to the torque screw or an out of specification hastalloy spring.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Common Device Name
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17466122
Report Number3004608878-2023-00141
Device Sequence Number7153325
Product Code HBL
UDI-Device Identifier10381780253549
UDI-Public10381780253549
Combination Product (Y/N)N
PMA/510(K) Number
K051440
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberA1114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/15/2023
Supplement Date Received by Manufacturer09/15/2023
Initial Report FDA Received Date08/04/2023
Supplement Report FDA Received Date09/21/2023
Date Device Manufactured04/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-