| Catalog Number |
386802 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.There were multiple lot numbers reported to be involved.The information for each additional lot number is as follows: d.4.Medical device lot #: 2174730.D.4.Medical device expiration date: 30-jun-2025.H.4.Device manufacture date: 23-jun-2022.D4.Medical device lot #: unknown.D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd cathena¿ safety iv catheter experienced leakage.The following information was provided by the initial reporter, translated from spanish to english: leakage at insertion side: "it is reported that after insertion of the catheter the insertion site is very dilated, which even causes the medication to leak out almost unnoticed.".
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 10-aug-2023.H.6.Investigation summary: our quality engineer inspected the 1 unused sample submitted for evaluation.The reported issues of phlebitis, catheter defective / damaged, leakage at insertion site, separation catheter from adapter were not confirmed upon inspection and testing of the samples.Analysis of the sample showed that there was no foreign matter present and no damages.The sample also underwent leakage testing, and no leakage was observed.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd cathena¿ safety iv catheter experienced leakage.The following information was provided by the initial reporter, translated from spanish to english: leakage at insertion side: "it is reported that after insertion of the catheter the insertion site is very dilated, which even causes the medication to leak out almost unnoticed.".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: imdrf annex b grid: b03, b05, b14.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd cathena¿ safety iv catheter experienced leakage.The following information was provided by the initial reporter, translated from spanish to english: leakage at insertion side: "it is reported that after insertion of the catheter the insertion site is very dilated, which even causes the medication to leak out almost unnoticed.".
|
| |
|