| Model Number |
9014 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 11/2024).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the drug-coated balloon was allegedly twisted upon inflation.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One fluoroscopic image was provided and reviewed.The image shows a balloon dilation within the superficial femoral artery.The mid balloon is not inflated with contract both proximal and distal to the non inflated portion.It demonstrates a partially twisted segment in the mid balloon.Therefore, the investigation is confirmed for the reported material twisting during inflation as twisting was noted on the balloon during dilation from the image review.A definitive root cause for the reported material twisting during inflation could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2024).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the drug-coated balloon allegedly twisted upon inflation.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|