| Catalog Number |
06P01-60 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a false repeat reactive alinity s hiv ag/ab combo result on a donor.The following results were provided: (b)(6) 2023 sid (b)(6)= 1.20 / 1.23 / 1.33 s/co; the confirmatory test was negative (hiv geenius).A new sample sid (b)(6) was negative on the alinity on (b)(6) 2023 = 0.96 s/co.No impact to donor management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of complaint data for the product and likely cause alinity s hiv ag/ab combo reagent lot 48221be00 identified normal complaint activity.Field data review showed that the repeat reactive rate observed for the complaint lot in blood banks meets the specificity performance claim of the package insert.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.No customer returns were available for evaluation.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lots.This review did not identify any non-conformances, potential non-conformances, or deviations.A review of labeling concluded that the issue is sufficiently addressed.The alinity s hiv ag/ab combo assay is performing as expected and no systemic issue or product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
The customer reported a false repeat reactive alinity s hiv ag/ab combo result on a donor.The following results were provided: 07apr2023 sid (b)(6) = 1.20 / 1.23 / 1.33 s/co; the confirmatory test was negative (hiv geenius).A new sample (sid (b)(6)) was negative on the alinity on 08apr2023 = 0.96 s/co.No impact to donor management was reported.
|
| |
|
Search Alerts/Recalls
|