ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE INSERTER; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Positioning Failure (1158); Fracture (1260); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: 42522100810 - articular surface medial congruent (mc) right 10 mm height use with tibia sizes e-f/cr femur sizes 8-11 - 65713314.G2: foreign country: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2023-00211.
 
Event or Problem Description
It was further reported that the inserter suffered a small break on the lower part that attaches to the tibial implant.A piece less than 1mm broke off and was retrieved.This caused the inserter to clip into the tibial implant in a position that was not deep enough to be secured therefore pushing the poly between the femoral component and tibial component in a crooked position which scarred the bottom of the poly.There were no surgical delays.The surgical technique was utilized.It is unknown how many times this instrument had been used.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Upon receiving additional information of the reported event, it was determined to be not reportable as there was no fracture or material separation seen on the device.The initial report should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE INSERTER
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17470229
Report Number0001822565-2023-02089
Device Sequence Number15234072
Product Code MBH
UDI-Device Identifier00889024656338
UDI-Public(01)00889024656338(10)65040196
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42529900100
Device Lot Number65040196
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2023
Supplement Date Received by Manufacturer12/07/2023
Initial Report FDA Received Date08/07/2023
Supplement Report FDA Received Date12/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient SexPrefer Not To Disclose
-
-