| Model Number |
DE2110X13B |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A ventilator was returned to a third-party service center for service.There was no harm or injury reported.During the evaluation of the device at third-party service center, it was observed that the device was malfunctioning.The device evaluation is still ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 other text : device not return.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported, a ventilator was returned to a third-party service center for service.There was no harm or injury reported.During the evaluation of the device at third-party service center, it was observed that the device was malfunctioning.The device's proportional valve (aecm) needs to be replaced to address the issue.H3 other text : third party service center.
|
| |
|