• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (PLEASANTON) COOLSCULPTING SYSTEM; DERMAL COOLING PACK/VACUUM/MASSAGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (PLEASANTON) COOLSCULPTING SYSTEM; DERMAL COOLING PACK/VACUUM/MASSAGER Back to Search Results
Catalog Number UNK CONTROL UNIT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperplasia (1906)
Date of Event 08/15/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Allergan aesthetics received a report of a male patient treated with coolsculpting to the lower abdomen on (b)(6) 2017 using coolcore, (b)(6) 2018 to the lower abdomen using coolcore applicator and on (b)(6) 2018 to the flanks using coolcourve+ applicator who developed paradoxical hyperplasia on lower abdomen and flanks on (b)(6) 2018 after treatment.
 
Additional Manufacturer Narrative
This is a historical record and reflects reports that were received by the company prior to april 2021.This mdr is associated with a parallel plate applicator, these applicators were discontinued and recalled in 2022.According to the coolsculpting user manual, under rare adverse events, paradoxical adipose hyperplasia (ph/pah) is characterized by a visibly enlarged tissue volume within the treatment area, which may develop two to five months after treatment.Surgical intervention may be required.Pah is not related to any coolsculpting device failure mode but it is observed more frequently with the parallel plate applicators.Pah is included in the risk management files of the device as an inherent risk to the use of cryolipolysis for localized fat reduction.
 
Event or Problem Description
Allergan aesthetics received a report of a male patient treated with coolsculpting to the lower abdomen on (b)(6) 2017 using coolcore, (b)(6) 2018 to the lower abdomen using coolcore applicator and on (b)(6) 2018 to the flanks using coolcourve+ applicator who developed paradoxical hyperplasia on lower abdomen and flanks on (b)(6) 2018 after treatment.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: updated serial number in the product information tab in parent record to ni.
 
Event or Problem Description
Allergan aesthetics received a report of a male patient treated with coolsculpting to the lower abdomen on (b)(6) 2017 using coolcore, (b)(6) 2018 to the lower abdomen using coolcore applicator and on (b)(6) 2018 to the flanks using coolcourve+ applicator who developed paradoxical hyperplasia on lower abdomen and flanks on (b)(6) 2018 after treatment.
 
Additional Manufacturer Narrative
Corrected data d1 : brand name.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOLSCULPTING SYSTEM
Common Device Name
DERMAL COOLING PACK/VACUUM/MASSAGER
Manufacturer (Section D)
ALLERGAN (PLEASANTON)
4410 rosewood drive
pleasanton CA 94588
Manufacturer (Section G)
ALLERGAN (PLEASANTON)
4410 rosewood drive
pleasanton CA 94588
Manufacturer Contact
terry ingram
4410 rosewood drive
pleasanton, CA 94588
9256214130
MDR Report Key17475008
Report Number3007215625-2023-11896
Device Sequence Number12616425
Product Code OOK
UDI-Device Identifier00816417020179
UDI-Public00816417020179
Combination Product (Y/N)N
PMA/510(K) Number
K160259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberUNK CONTROL UNIT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/08/2018
Supplement Date Received by Manufacturer09/15/2023
10/12/2023
Initial Report FDA Received Date08/07/2023
Supplement Report FDA Received Date09/15/2023
11/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1347-2022, Z-1348-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight94 KG
-
-