| Model Number |
3560 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Defective Device (2588); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the burr got stuck on wire.A rotawire guidewire was selected for use and was dressed with a wet compress upon unpacking.The rotawire guidewire and rotapro device were used together during the procedure.During pullback of the devices, it was difficult to track the rotapro burr on the guidewire and the wire appeared blunt.The rotawire was removed manually and in dynaglide mode.The rotawire guidewire was removed intact and the procedure was able to be completed with the same rotapro device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter telephone: (b)(6).Device evaluated by manufacturer: the rotawire: ic device was returned for investigation.Visual and microscopic investigation confirmed the spring tip was observed to be bent and stretched.
|
| |
|
Event or Problem Description
|
|
It was reported that the burr got stuck on wire.A rotawire guidewire was selected for use and was dressed with a wet compress upon unpacking.The rotawire guidewire and rotapro device were used together during the procedure.During pullback of the devices, it was difficult to track the rotapro burr on the guidewire and the wire appeared blunt.The rotawire was removed manually and in dynaglide mode.The rotawire guidewire was removed intact and the procedure was able to be completed with the same rotapro device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|