STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080617
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the insulation has been compromised.
 
Event or Problem Description
It was reported that the insulation has been compromised.
 
Additional Manufacturer Narrative
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: it was reported by the customer that the laparoscopic grasping forceps had the black insulation peel off inside the patient.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be gouging or scraping of the insulation material, which may have occurred during use and/or sterilization causing the material to begin to peel.The lot number is unknown, therefore the manufacture date cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM PEEK MONOPOLAR HANDLE
Common Device Name
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key17476444
Report Number0002936485-2023-00761
Device Sequence Number7366509
Product Code OCZ
UDI-Device Identifier07613327058567
UDI-Public07613327058567
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K973259
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0250080617
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/11/2023
Supplement Date Received by Manufacturer07/11/2023
Initial Report FDA Received Date08/07/2023
Supplement Report FDA Received Date07/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1