• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC EXPEDIUM SPINE SYSTEM COMPRESSOR, PARALLEL ACTION; PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC EXPEDIUM SPINE SYSTEM COMPRESSOR, PARALLEL ACTION; PROBE Back to Search Results
Catalog Number 279702070
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2023
Type of Reportable Event Malfunction
Event or Problem Description
Device report from depuy synthes reports an event in japan as follows: it was reported that this was an unknown surgery for ais (type2) performed on (b)(6) 2023.After setting rods, the surgeon used the compressor for leveling the vertebral body, but the compressor was difficult to apply force therefore competitor¿s compressor which the surgeon has been familiar to use for setting the verse system and altalyne rods was used.A phenomenon occurred in which the verse system and altalyne rods moved even though the surgeon applied a strong compression and the assistant surgeon made the final tightening.Therefore, the surgeon become enraged that the verse system might be systemically problematic in cases such as scoliosis, which requires strong correction.A competitor's device was used in the distraction procedure too and the final tightening was done, but the verse system could not hold the tightened position after the final tightening.The surgeon pointed out the shortcomings of the verse system.To prevent the implants from moving from the position set by the final tightening, the surgeon distributed the stress so that it was not concentrated in one place and managed to complete the procedure.The surgery was completed successfully with no surgical delay.No further information is available.This report is for one (1) expedium spine system compressor, parallel action this is report 1 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H3, h4, h6: without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPEDIUM SPINE SYSTEM COMPRESSOR, PARALLEL ACTION
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key17476604
Report Number1526439-2023-01441
Device Sequence Number7270149
Product Code HXB
UDI-Device Identifier10705034198460
UDI-Public(01)10705034198460
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number279702070
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/24/2023
Initial Report FDA Received Date08/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - RODS: SPINE.; UNK - SCREWS: EXPEDIUM.
Patient SexUnknown
-
-