| Model Number |
LPG1614 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, in a laparoscopic inguinal hernia repair, as the mesh was being introduced through the trocar, the valve seal completely came off.The trocar was attempted to be fixed.The surgeon tried re-fitting the outer grey seal.Due to rapid and continued loss of abdominal pressure, the balloon was deflated, removed, and the device was replaced.The mesh which was being used unintendedly got stuck to the tissue.The length of the procedure increased while the surgeon detached the mesh from this unintended tissue.The mesh was still able to be used, and was successfully implanted.No device component fell into the patient¿s cavity.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant product: oms-t10sboms-t10sb trocar balloon o armatures10, lot# p3a0237.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|