| Catalog Number |
5F050603CS |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 04/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure in the superficial femoral artery, a piece of delivery system was allegedly broke off when they were trying to remove the delivery system.It was further reported that they were not able to snare the piece that was left behind to remove it and unable to get the leftover piece that way and the patient had to be taken to surgery to remove it.The current status of the patient is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure in the superficial femoral artery, a piece of delivery system was allegedly broke off when they were trying to remove the delivery system.It was further reported that they were not able to snare the piece that was left behind to remove it.Reportedly, they were unable to get the leftover piece that way, and the patient had to be taken to surgery to remove it.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation and no x-ray images were provided which leads to inconclusive evaluation result.Based on the information available the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instruction for use states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the instruction for use states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' under materials required the instruction for use states: '5f (1.67 mm) or larger introducer sheath (¿); 0.014-inch (0.36 mm) - 0.035-inch (0.89 mm) diameter guidewire'.Regarding insertion and removal difficulty of the delivery system the instruction for use states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' the instruction for use further state: 'examine the stent system for any damage.If it is suspected that the sterility or performance of the stent system has been compromised, the device should not be used.' holding and handling of the system throughout deployment was found sufficiently described.H10: d4 (expiration date: 04/2025), g3.H11: h6 (method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|