| Model Number |
BTB TIGHTROPE W/ DEPLOYING SUTURE |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 7/11/2023, it was reported by an arthrex employee via email that there was strong resistance when pulling the wire snare from an ar-1588btb-j btb tightrope.The case was completed using a new product.This was discovered on the back table while prepping the graft during a procedure on (b)(6) 2023.Additional information requested.Additional information provided 7/11/2023: the procedure performed was an acl repair.The case was completed by using a new ar-1588btb-j.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h1 to n/a.Arthrex previously submitted this event as a reportable malfunction out of an abundance of caution.Upon further review and evaluation of associated risk documentation, it has been determined that this event does not meet the criteria of a reportable event under 21 cfr 803.Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further review and evaluation of associated risk documentation, it has been determined that this event does meet the criteria of a reportable event under 21 cfr 803.
|
| |
|