• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BTB TIGHTROPE W/ DEPLOYING SUTURE; SMOOTH FIXATION PIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BTB TIGHTROPE W/ DEPLOYING SUTURE; SMOOTH FIXATION PIN Back to Search Results
Model Number BTB TIGHTROPE W/ DEPLOYING SUTURE
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 7/11/2023, it was reported by an arthrex employee via email that there was strong resistance when pulling the wire snare from an ar-1588btb-j btb tightrope.The case was completed using a new product.This was discovered on the back table while prepping the graft during a procedure on (b)(6) 2023.Additional information requested.Additional information provided 7/11/2023: the procedure performed was an acl repair.The case was completed by using a new ar-1588btb-j.
 
Additional Manufacturer Narrative
Correction: h1 to n/a.Arthrex previously submitted this event as a reportable malfunction out of an abundance of caution.Upon further review and evaluation of associated risk documentation, it has been determined that this event does not meet the criteria of a reportable event under 21 cfr 803.Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
 
Additional Manufacturer Narrative
Upon further review and evaluation of associated risk documentation, it has been determined that this event does meet the criteria of a reportable event under 21 cfr 803.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BTB TIGHTROPE W/ DEPLOYING SUTURE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17479649
Report Number1220246-2023-07446
Device Sequence Number2598183
Product Code HTY
UDI-Device Identifier00888867095366
UDI-Public00888867095366
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112990
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBTB TIGHTROPE W/ DEPLOYING SUTURE
Device Catalogue NumberAR-1588BTB-J
Device Lot Number14953930
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2023
Supplement Date Received by Manufacturer01/16/2025
Not provided
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date02/10/2025
02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-