• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 40MM-STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 40MM-STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 412.214S
Medical Device Problem Codes Device-Device Incompatibility (2919); Migration (4003)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on an unknown date, the patient experienced fns (femoral neck screw) cutout.During a removal procedure, the fns locking screw was cold welded into the plate.The patient had an additional fracture during removal of the screw.The procedure was completed successfully, but took longer than expected according to the surgeon.The patient outcome was satisfactory.This report involves one 5.0mm ti locking scr slf-tpng w/t25 stardrive 40mm-sterile.This is report 3 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part number: 412.214s-us, lot number:349p997, manufacturing site: mezzovico, release to warehouse date: 15 sep 2021, expiration date: 31 aug 2031.A manufacturing record evaluation was performed for the sterile finished lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 40MM-STERILE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK MEZZOVICO (CH)
mezzovico
via cavazz 5 6805
SZ   6805
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17479969
Report Number8030965-2023-09976
Device Sequence Number16767652
Product Code KTT
UDI-Device Identifier10886982275417
UDI-Public(01)10886982275417
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K000682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number412.214S
Device Lot Number349P997
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/24/2023
Initial Report FDA Received Date08/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.0 TI LCKNG SCR SLF-TPNG T25 SD 40-SILE; ANTIROTATION SCR FEM NECK 95 LNTH -S; BOLT FOR FEMORAL NECK SYS 90 LENGTH-SILE; FEMORAL NECK SYSTEM PL 2 HOLE - STERILE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-