GYRUS ACMI, INC. ULTRA-CATCH NT (1.8 FR X 115 CM) STONE RETRIEVAL DEVICE Back to Search Results
Model Number NT4W18115
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported to olympus upon opening the box and checking the actual product, the basket wire was cute.The therapeutic transurethral lithotripsy procedure was completed with a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The subject device is expected to be returned to olympus for further evaluation and testing.The investigation is in process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.If additional information becomes available, this report will be supplemented accordingly.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Updated fields: h4, h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was manufactured in march 2023.As the device was not returned, a physical device inspection could not be performed.As a result, the complaint could not be confirmed, and potential causes could not be determined the suggested event is detectable and preventable by handling the scope in accordance with the following section of the instructions for use: "prior to use, inspect the device for any visible damage such as kinks or broken stone dislodger wires; do not use the product should the product or its container be received damaged.Contact olympus for authorization to return the product for reinspection and/or replacement." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA-CATCH NT (1.8 FR X 115 CM) STONE RETRIEVAL DEVICE
Common Device Name
STONE RETRIEVAL DEVICE
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17480463
Report Number3011050570-2023-00111
Device Sequence Number21076540
Product Code FGO
UDI-Device Identifier00821925027725
UDI-Public00821925027725
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNT4W18115
Device Lot NumberKR276868
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2023
Supplement Date Received by Manufacturer09/22/2023
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date09/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-