• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1.5MM X 4CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1.5MM X 4CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035431540
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 07/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3 other text : the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the left middle cerebral artery m2 aneurysm case while delivery of the subject coil, resistance was encountered in the microcatheter.On withdrawal of the subject coil, thrombus was generated in the microcatheter.The thrombus was aspirated from microcatheter.The physician replaced the subject coil with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.During visual/microscopic inspection, the coil delivery wire was noted to be kinked/bent.The main coil was noted to be stretched.The introducer sheath and catheter were not returned.Functional testing was unable to perform due to damage.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications as the catheter was not returned along with the introducer sheath.Additional information provided by the customer indicated that no damage was noted to the packaging prior to opening the packaging.The device was prepared for use as per the directions for use.Continues flush was maintained throughout the procedure.Resistance was encountered in the microcatheter.The coil was withdrawn and found thrombus generated in microcatheter.An assignable cause of procedural factors will be assigned to the as reported 'coil in catheter friction' as these defects appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Based upon medical review, the harm observed in this complaint is anticipated in nature as per the device risk assessment.The reported 'patient vessel thrombosis is a known and anticipated complication to these types of procedures and patient condition and is listed as such in the device directions for use.Therefore, a probable cause of anticipated procedural complication was assigned to this event.An assignable cause of procedural factors will be assigned to the as analyzed 'coil delivery wire kinked/bent' and 'main coil stretched' as these defects appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.
 
Event or Problem Description
It was reported that during the left middle cerebral artery m2 aneurysm case while delivery of the subject coil, resistance was encountered in the microcatheter.On withdrawal of the subject coil, thrombus was generated in the microcatheter.The thrombus was aspirated from microcatheter.The physician replaced the subject coil with a new device and continued the procedure without clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET HELICAL NANO 1.5MM X 4CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key17480989
Report Number3008881809-2023-00404
Device Sequence Number15678967
Product Code HCG
UDI-Device Identifier04546540698056
UDI-Public04546540698056
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/08/2025
Device Catalogue NumberM0035431540
Device Lot Number23692073
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2023
Supplement Date Received by Manufacturer08/31/2023
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER (STRYKER); SYNCHRO GUIDEWIRE (STRYKER)
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient RaceAsian
-
-