| Model Number |
CD2231-40Q |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Interrogate (1332)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
| |
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and device was not able to interrogate.The device was explanted and replaced.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with lithium (li) cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|
| |
|