| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was unable to duplicate the reported issue.The issue was verified through review of the software logs.The wire harness, battery pins and internal/external screws were replaced.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly lost power.In this state the device would not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
|
| |
|