• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.A 8 x 80 x 120 epic vascular stent was selected for use.However, after unpacking, it was noticed that there was a space at the tip of the stent.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
A2 - age at time of event: 18 years or older.E1 - initial reporter phone: (b)(6).
 
Event or Problem Description
It was reported that stent damage occurred.A 8 x 80 x 120 epic vascular stent was selected for use.However, after unpacking, it was noticed that there was a space at the tip of the stent.The procedure was completed with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
A2 - age at time of event: 18 years or older.E1 - initial reporter phone: (b)(6) epic vascular 8x80x120 was received for analysis.The device was received with the stent partially deployed on the delivery system.No issues were noted with the stent.Blood and possible body matter was identified on the device.A visual examination identified no issues with the tip of the device.A visual and tactile examination found the sheath of the device to be damaged.This type of damage is consistent with excessive force being applied to the device.A visual examination identified no issues or damage to the handle of the device.The safety lock was in place on the rack of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17482572
Report Number2124215-2023-39425
Device Sequence Number15239921
Product Code NIO
UDI-Device Identifier08714729805724
UDI-Public08714729805724
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0030326335
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/19/2023
Supplement Date Received by Manufacturer08/22/2023
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-