Model Number EGIA30AMT
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Codes Laceration(s) (1946); Tissue Breakdown (2681)
Date of Event 07/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: sigadaptstnd, sig power sigadaptstnd linear adapter.Serial#: (b)(6).Sigphandle, sig power sigphandle handle (lot#: unknown) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but three photos were available for evaluation.Visual inspection noted the first returned image contained a view of the operating room.Two handles, clamshells, and adapters could be seen inserted into the patient.A monitor in the background of the image contained a view of two reloads adjacent to the tissue.The second returned image contained a view of a manual retract tool.The tool was noted to be broken and separated into its component parts.The third returned image contained a view of a reload with tissue clamped in the jaws.The knife blade was noted to be fully advanced and the staples that were not inserted into the tissue were noted to be properly formed.It was reported that the jaws of the device remain closed on tissue.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, in a laparoscopic appendectomy, while in the appendix, the jaws of the device locked on tissue.The jaws would not open.A new power stapler and a new trocar were opened to release the tissue.The surgeon stapled a little further down to release the tissue and the reload.The incision was also extended.
 
Additional Manufacturer Narrative
Additional information: b1, b2, b5, g3, h1, h6 (imf code) new information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a serious injury.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, in a laparoscopic appendectomy, while in the appendix, the jaws of the device locked on tissue.The jaws would not open.A new power stapler was opened and a new trocar had to be added to pass the new stapler to release the tissue.The incision had to be extended to redo the passage for the new 12mm trocar.The surgeon stapled a little further down to release the tissue and the reload.
 
Additional Manufacturer Narrative
New information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a malfunction.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the reload was returned loaded on an adapter with the jaws closed.Gapping was observed at the attachment site of the reload to the adapter.The reload was fully fired and the knife bar was fully advanced.Staples were present between the jaws.No tissue was present between the jaws.The laminates were noted to be herniated and sticking out of the reload.Functional testing noted that the reload was unable to be removed from the adapter.It was reported that the jaws of the device locked on tissue.The reported issue was confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition of herniated laminates.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17483555
Report Number1219930-2023-03372
Device Sequence Number15153261
Product Code GDW
UDI-Device Identifier10884523003062
UDI-Public10884523003062
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEGIA30AMT
Device Catalogue NumberEGIA30AMT
Device Lot NumberN2G0710Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2023
Supplement Date Received by Manufacturer07/26/2023
08/17/2023
04/26/2024
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date08/16/2023
09/08/2023
05/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H10
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-