| Catalog Number |
826633 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported: "five hours after the microsensor was implanted into the patient, there was no icp readings showed on the icp monitor.Then the physician used this icp monitor to connect with microsensor of other patient, it could show the icp readings normally.Therefore, the problem was with the microsensor.Then, the physician changed with another of the same device.".
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The microsensor (id 826633) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: d9, g3, g6, h2, h3, h6, h11.Device history record (dhr): there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis: catheter mashed 6.5cm from connector.Catheter internal wires are broken (x-ray).Icp express read "no transducer detected".No further testing possible.Root cause analysis: the returned device presents with the catheter mashed 6.5cm from connector and internal wires broken.The complaint has been confirmed with the probable cause being related to mishandling of the device.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|