INTEGRA LIFESCIENCES MANSFIELD ICP SENSOR CATHETER KIT; ICP MICROSENSORS Back to Search Results
Catalog Number 826633
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported: "five hours after the microsensor was implanted into the patient, there was no icp readings showed on the icp monitor.Then the physician used this icp monitor to connect with microsensor of other patient, it could show the icp readings normally.Therefore, the problem was with the microsensor.Then, the physician changed with another of the same device.".
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The microsensor (id 826633) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Additional Manufacturer Narrative
Updated fields: d9, g3, g6, h2, h3, h6, h11.Device history record (dhr): there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis: catheter mashed 6.5cm from connector.Catheter internal wires are broken (x-ray).Icp express read "no transducer detected".No further testing possible.Root cause analysis: the returned device presents with the catheter mashed 6.5cm from connector and internal wires broken.The complaint has been confirmed with the probable cause being related to mishandling of the device.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICP SENSOR CATHETER KIT
Common Device Name
ICP MICROSENSORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key17483967
Report Number3014334038-2023-00132
Device Sequence Number6770495
Product Code GWM
UDI-Device Identifier10381780520436
UDI-Public10381780520436
Combination Product (Y/N)N
PMA/510(K) Number
K991222
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number826633
Device Lot Number6396430
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/26/2023
Supplement Date Received by Manufacturer10/03/2023
09/07/2024
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date10/13/2023
09/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-