Model Number VICM5_12.6
Medical Device Problem Codes Misfocusing (1401); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Blurred Vision (2137)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
 
Event or Problem Description
The reporter indicated that a 12.6mm vticm5_12.6 implantable collamer lens of -13.50 diopter was implanted into the patient's right eye (od) on (b)(6) 2023.Patients vision after surgery was 20/50 +1 with vault around 1000.After 2 weeks with no visual improvement, the lens was exchanged for a toric lens of shorter length and different power.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
B5- per updated information the problem resolved.Claim # (b)(4).
 
Additional Manufacturer Narrative
Corrected data: b5- the reporter indicated the surgeon implanted a 12.6mm vicm5_12.6 implantable collamer lens of diopter -13.50 into the patient's right eye (od) on (b)(6) 2023.The patient experienced an excessive vault and refractive surprise.Reportedly "no visual improvement" (va 20/50).Higher vaulting was also noted".On (b)(6) 2023 the lens was exchanged with a toric model, shorter length and the problem was resolved.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key17484564
Report Number2023826-2023-03356
Device Sequence Number7383194
Product Code MTA
UDI-Device Identifier00841542119252
UDI-Public00841542119252
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberVICM5_12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2023
Supplement Date Received by Manufacturer08/14/2023
03/26/2024
Initial Report FDA Received Date08/08/2023
Supplement Report FDA Received Date09/08/2023
03/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-