The device was returned and evaluated, and the customer¿s allegation of tip knife wire was cut was confirmed.Device evaluation found that the cutting wire was broken, scorched and melted.The knife wire was deformed.The outer diameter of the cutting wire was measured.The result indicated no abnormalities.The length of the coated portion of the cutting wire, and the cutting wire itself presented no abnormalities.There were no missing parts in the subject device.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than a year since the subject device was manufactured.The directional defect of the knife wire could not be confirmed.A definitive root cause cannot be identified.However, based on the results of the investigation, it is likely that the following led to the malfunction: the knife wire broke due to discharge during energization, and the knife wire did not point in the intended direction.Based on the results of the actual product confirmation and past investigations, it is assumed that the knife wire was broken by the following mechanism.1.The device did not protrude enough from the endoscope until the rear end of the cutting wire was in the field of view.2.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were close to each other.3.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.4.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.This issue is addressed in the instructions for use (ifu): this instruction manual contains the following information.¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿ when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Olympus will continue to monitor the field performance of this device.
|