ETHICON INC. SFX SYM PDS+ UNI VIO 18IN 1 S/A CTXB; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number SXPP1A300
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Trade name - irgacare® active ingredient(s) ¿ triclosan dosage form ¿ suture/solid/parenteral strength ¿ = 2360 g/m.H3 investigation summary: the product was returned for evaluation.Visual inspection was conducted on the returned device.The returned sample revealed that it was received, one open sample that pertain to the product code sxpp1a300.During the visual assessment of the needle suture, the suture was examined, body fluids and damage at some barbs were noted and both sides of the fixation tab were broken.It should be noted that a 100% inspection is performed via the automotive vision system to ensure the tab/stopper, barbs and core is present and complete during manufacturing.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The instructions for use contain the following caution: to seat the fixation tab; pull the device through the tissue to gently seat the fixation tab against the tissue.The fixation tab should be seated above the tissue plane and be visible.Do not exert additional force on the fixation tab or device.Moving toward the apex of the incision, take a pass in the intact tissue perpendicular to the initial pass to lock the stitch.Multiple passes are acceptable.Pull gently on the device to take up any slack.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and barbed suture was used.It was reported by a sales rep that, before use it was found that the fixation tab was already broken.During the visual assessment of the returned needle suture, the suture was examined, body fluids and damage at some barbs were noted and both sides of the fixation tab were broken.No adverse patient consequences were reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFX SYM PDS+ UNI VIO 18IN 1 S/A CTXB
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key17486702
Report Number2210968-2023-05709
Device Sequence Number12730735
Product Code NEW
UDI-Device Identifier10705031219359
UDI-Public10705031219359
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K113004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1A300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2023
Initial Report FDA Received Date08/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-