ZIMMER DENTAL 3I T3 TAPERED IMPLANT 6/5 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number BOPT6511
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A4: weight unknown / not provided.E1: last/given name unknown / not provided.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the internal components of the implant at tooth location #19 was stripped.The procedure was completed with another implant.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 180.H6: investigation conclusions code was added: 4307.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use.The implant was observed severely damaged on its side, and was drilled into it which damaged the implant's internal drive feature.At the bottom of the implant¿s internal threads, it appears that a screw was possible drilled out leaving a fragment behind.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number 2021111615.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2021111615 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are missing or confusing instructions for use, or the torque or speed applied during placement/seating exceeded recommended value, clinician inadvertently selects an instrument with the wrong drive feature to place/seat the implant.Therefore, based on the available information, device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 TAPERED IMPLANT 6/5 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17486774
Report Number0001038806-2023-01487
Device Sequence Number7376360
Product Code DZE
UDI-Device Identifier00844868029861
UDI-Public(01)00844868029861(17)261221(10)2021111615
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberBOPT6511
Device Lot Number2021111615
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/12/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date01/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age31 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-