This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 180.H6: investigation conclusions code was added: 4307.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use.The implant was observed severely damaged on its side, and was drilled into it which damaged the implant's internal drive feature.At the bottom of the implant¿s internal threads, it appears that a screw was possible drilled out leaving a fragment behind.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number 2021111615.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2021111615 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are missing or confusing instructions for use, or the torque or speed applied during placement/seating exceeded recommended value, clinician inadvertently selects an instrument with the wrong drive feature to place/seat the implant.Therefore, based on the available information, device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|