• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HICKMAN CV CATHETER, DUAL-LUMEN, 9F; CHRONIC CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HICKMAN CV CATHETER, DUAL-LUMEN, 9F; CHRONIC CATHETERS Back to Search Results
Catalog Number 0600600
Medical Device Problem Codes Burst Container or Vessel (1074); Fluid/Blood Leak (1250); Difficult to Flush (1251); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiry date: 07/2026).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately six months post chronic catheter placement, a cut was allegedly found at the lumen of the catheter.It was further reported that the catheter was allegedly not able to be flushed, and blood allegedly leaked from the catheter.Reportedly, the catheter was removed.There was no reported patient injury.
 
Event or Problem Description
It was reported that approximately six months post chronic catheter placement, a cut was allegedly found at the lumen of the catheter.It was further reported that the catheter was allegedly not able to be flushed, and blood allegedly leaked from the catheter.Reportedly, the catheter was removed.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one 9fr hickman d/l catheter was received for evaluation and one photo was provided for review.Visual, microscopic, tactile and functional evaluations were performed.A s-shaped split was noted on the catheter body, proximal to the bifurcate.Upon infusion, water was observed exiting the s-shaped split on the catheter body.The photo review also shows a s-shaped split in the catheter.Therefore, the investigation is confirmed for the reported fluid leak and the identified burst issue.The investigation is unconfirmed for the reported fracture issue as no objective evidence of fracture was noted.However, the investigation is inconclusive for the reported difficult to flush issue as the exact circumstance at the time of the event reported was unknown.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN CV CATHETER, DUAL-LUMEN, 9F
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17486998
Report Number3006260740-2023-03409
Device Sequence Number7881360
Product Code LJS
UDI-Device Identifier00801741051890
UDI-Public(01)00801741051890
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0600600
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/17/2023
Supplement Date Received by Manufacturer12/19/2023
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-