• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 5 RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 5 RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71423235
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a tka had been performed on (b)(6) 2022, the patient had an infected knee.A washout and revision surgery were performed on (b)(6) 2023 to address this adverse event.In addition, a legion ps np fem sz 5 rt and a gii p/s all poly tib sz3 15mm rt were replaced with s+n devices.
 
Additional Manufacturer Narrative
H3, h6: the devices were not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have definitively contributed to the event.Reportedly the patient underwent a washout and revision due to an unspecified infected knee (approximately) 9 months post total knee arthroplasty surgery.No blood or wound culture results were provided.The legion ps np fem sz 5 rt and a gii p/s all poly tib sz3 15mm rt were replaced with smith and nephew components.The patient impact beyond the reported knee infection and subsequent revision cannot be determined.The patient¿s current health status is unknown.No further clinical/medical assessment can be rendered.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.For the femoral component, a review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.For the tibial baseplate, a review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that acute post-surgical wound infection has been identified in possible adverse effects.Active, local infection or previous intra-articular infections are identified in contraindications for total knee replacement.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.A review of the sterilization records revealed the batches were sterilized within normal parameters.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include loss of sterility during procedure, post-operative healing issue, injury, patient condition and/or medical history.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS NP FEM SZ 5 RT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17487143
Report Number1020279-2023-01570
Device Sequence Number7253009
Product Code JWH
UDI-Device Identifier00885556030400
UDI-Public00885556030400
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K060742
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71423235
Device Lot Number22BM06553
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2023
Supplement Date Received by Manufacturer09/05/2023
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date09/06/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GII P/S ALL POLY TIB SZ3 15MM RT (71420438)
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-