Model Number CC60WF
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, lens was explanted with clinical reason as lens was cracked/fragmented.The iol was exchanged for an unknown atiol following the initial implant procedure.Additional information has been requested.
 
Event or Problem Description
Additional information has been requested and received stating that lens was exchanged with a non company lens, new information received states lens was also dislocated due to poor zonules.
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The product investigation could not identify a root cause for the reported complaint.The product has not been received to evaluate.Information was provided that the lens was also dislocated due to poor zonules.Associated products were not provided.It is unknown if qualified products were used.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.A company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.During lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.Follow up attempts were made.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17487642
Report Number1119421-2023-01424
Device Sequence Number15424185
Product Code HQL
UDI-Device Identifier00380652396974
UDI-Public00380652396974
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2026
Device Model NumberCC60WF
Device Lot Number15494922
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2023
Supplement Date Received by Manufacturer08/21/2023
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date09/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/20/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age32 YR
Patient SexMale
-
-