| Model Number |
RAD-SC30-175 |
| Medical Device Problem Code |
Device Contamination with Body Fluid (2317)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The oad was returned for analysis.There was no damage or abnormalities observed with the driveshaft or handle assembly that could have contributed to the reported event.There was no blood observed within the saline sheath or saline line assembly.Fluid flow testing through the oad and saline sheath did not identify any irregularities.When tested, the oad functioned as intended.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
| |
|
Event or Problem Description
|
|
A stealth 360 peripheral orbital atherectomy device (oad) was used for six treatments.After the treatments, it was noticed that the amount of viperslide in the bag was not decreasing despite a lot of flushing.Blood was observed in the saline tubing.The treatment was completed with no further issues.
|
| |
|