| Catalog Number |
8065752062 |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported cutter was in bad condition and actuation failure occurred during vitrectomy surgery.The condition of aspiration was unknown.The surgery was completed after replacing the product with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
One opened probe was received with a tip protector, in a tray with other items.The sample was visually inspected and found to be non-conforming with orange/brown and clear foreign material on the port face.The sample was then functionally tested for actuation and was found to be non-conforming.The probe sample was disassembled, and the components inspected.Nominal wear was observed on the inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Wear marks were observed at one location along the inner cutter.The sample was retested for actuation with the probe driver and was able to actuate.The initial actuation test failed due to an interference within the probe and once the interference (needle assembly) was removed the probe was able to actuate.The needle holder and retainer were then disassembled.All components were inspected and found to be conforming.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification (rfid) tag indicates that the product was processed and released according to the product¿s acceptance criteria.The complaint evaluation confirms that the probe sample had an actuation failure.The exact cause of the actuation failure cannot be determined from the evaluation performed.The exact root cause of the actuation failure could not be determined, therefore, no specific action with regard to this complaint was taken by the manufacturing site.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints will be continued to be reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|