| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - medical product: ng rot.Hinge knee tib plt sz4 catalog # 00-5880-004-00 lot # 63901409r, art surface 12mm fem sz e catalog # 00-5880-050-12 lot # 64834810, stem implant 12mmdx145mm catalog # 00-5988-010-12 lot # 65479568, stem implant 14mmdx145mm catalog # 00-5988-010-14 lot # 64509180, 15mm tib aug blk sz4 rlat/lmed catalog # 00-5448-004-28 lot # 63923034, 36 deg ele prov liner set 36mm catalog # 00-5490-035-02 lot # 64264990, 5mm distal fem aug sz e catalog # 00-5490-035-10 lot # 64575595.G2 : japan.H3: customer has indicated that the product will not be returned because it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported patient underwent a revision procedure nine months post implantation due to hinge post extension backed out and came off.Attempts to obtain additional information have been made; however, no more is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records provided.Device was not returned.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.X-ray review indicates a left knee arthroplasty is present with overall anatomic alignment.The hinge post is displaced and separated from the implant with implant displacement anteroinferiorly.Radiolucency is noted along the distal aspect of the femoral implant without evidence of implant loosening.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|