Model Number N/A
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Fall (1848); Pain (1994); Swelling/ Edema (4577)
Date of Event 05/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2023-00201.D10- medical product articular surface (mc) left 12 mm height; item# 42512100312; lot# 64286787.G2- australia.H3- customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient was revised approximately one year three and a half months post implantation due to instability.The patient had both pain and swelling prior to the revision.The surgeon doesn't believe that the original components (or failure of) contributed to the fall(s) of the patient.Attempts to obtain additional information have been made; however, no more information is available.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.No devices or photos were received; therefore, the condition of the components is unknown.X-ray review by third party hcp states that punctate metallic foreign body medial compartment.Large ossific densities in the region of the suprapatellar bursa on the lateral film.Mildly hyperdense suprapatellar joint effusion is of uncertain clinical significance.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.Root cause cannot be determined.This complaint cannot be confirmed for pain and instability however swelling can be confirmed via x-rays provided.H6: component code: proposed component (annex g) code is: - mechanical (g04) - femur.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR TRABECULAR METAL (CR) STANDARD POROUS
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17490796
Report Number0001822565-2023-02073
Device Sequence Number20801391
Product Code MBH
UDI-Device Identifier00889024230712
UDI-Public(01)00889024230712(17)270930(10)63773065
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K122745
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502805801
Device Lot Number63773065
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Supplement Date Received by Manufacturer11/03/2023
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date11/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight86 KG
-
-