| Model Number |
26159EHW |
| Medical Device Problem Codes |
Crack (1135); Material Fragmentation (1261); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
Tiny metal fragments (powder-like) were dislodged from the scissors when it was inserted into the patient's uterus via single-flow operating sheath.One larger fragment attempted to be removed but washed out when trying to transfer to the specimen container.
|
| |
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Visual and microscopical examination revealed that the cutting surfaces show the usual signs of wear considering their age (manufacturing date - august 2018).A closer inspection of the blades does not reveal any defects.The function of the blades was tested and is given.Since the article is already from 2018, defects / product issues on the cutting blades would have been noticed earlier.It cannot be confirmed that the foreign particles are material from the shears.The ifu refers to the service life of the product and regular testing.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|