Zimmer biomet complaint number (b)(4).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost non-existent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Summary investigation.
|
|
It was reported that the implant at tooth site # 47 was removed due to peri-implantitis.At uncovering of the upper 4-5 threads, the implant stood approximately 3-4 mm out of the bone (while implant was at bone level) and doctor performed bone regeneration.The doctor performed uncovering at tooth site # 47 and noticed a lingual pocket of 8 mm, implant had to be removed.Bone regeneration was performed at tooth site # 46.
|