| Catalog Number |
0620040610 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was inlet gas pressure issues during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was inlet gas pressure issues during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: gas issue says bottle gas.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be a loosely connected gas adapter and/or the accessories used along with the unit at the time of the event.Manufacture date is not known.
|
| |
|