| Catalog Number |
520A1299US |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device needed repair for not cycling.Patient involvement unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: additional information is provided in h.2., h.3., and h.6.Functional testing did not confirm the customers complaint.The root cause for this event is unknown.The product is beyond a year from manufacture date and there was no indication of a manufacturing defect during the investigation, so a dhr review was not performed.The service history review found the previous service replaced the rolling diaphragm which is related to the complaint.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.D.4.Full udi number: (b)(4).
|
| |
|