|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, as well as corrections to d4 and g2.H4: based on the 3 digit lot number provided, the manufacturing date of the device was in the month of november 2022 but a specific date could not be identified.Three attempts were performed to obtain additional information, but no response was received from the customer.The device was not returned as it was discarded; therefore, the reported phenomenon or condition of the device could not be confirmed.However, it is possible to infer the cause from the results of past similar investigations.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of past investigations, it is likely that the probe broke due to contact with a surgical instrument or the non-insulated area of grasping section.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.During the output activation in seal & cut mode, the probe came into contact with hard tissue, metal objects or surgical instruments.This resulted in scratches.2.A force to activate the output in seal & cut mode, or a force to grasp the body tissue was applied to the probe.Therefore, cracks were branching from the scratches on the probe.3.A force was applied to the probe causing it to break.Or 1.Grasping section was closed without grasping anything while the output was activated in seal & cut mode (this includes after tissue resection).This caused the tissue pad to wear out.2.The tissue pad was worn out.This has resulted in the non-insulated area and the distal end of the probe coming into contact with each other.3.The output was activated in seal & cut mode in state of above description.This has resulted in scratches (contact marks) on the two distal ends of the probe and on the grasping section.4.A force was applied to the probe to activate the output in the seal & cut mode, or a force was applied to grasp the tissue.This resulted in cracks branching at the scratched area.5.A force was applied to the probe causing it to break.The following are the instructions for use which state: "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.The thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, other instruments, or forceps, and others.Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/split/protruding/partial separating of the tissue pad.In turn, the probe tip may break before displaying an error window or generating an alarm tone.During output, do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments.Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities, which may lead to burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure".Olympus will continue to monitor field performance for this device.
|