| Catalog Number |
04906969001 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter complained of a software issue with a cobas u 411 urine analyzer.The customer alleged the sequence number of the patient¿s sample was changed leading to a mismatch in patient results.The specific results were not provided.The customer reviewed the urine sediment from the test and performed a visual verification where the issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
The u 411 analyzer serial number was (b)(6).The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The field application specialist (fas) confirmed that on the day of the event, there was an issue with the result sequence and the result transmission.No data or actual results were provided for the investigation.Based on other information provided, the investigation noted there were some instrument issues when placing the test strip onto the test strip tray.The field service engineer (fse) replaced the printed circuit boards (pcb) for the test strip detector and the strip check.After the service visit, the customer had no further issues.As no specific data was provided for the investigation, the cause of the event could not be determined.
|
| |
|