• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS U 411 URINE ANALYZER; AUTOMATED URINALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS U 411 URINE ANALYZER; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 04906969001
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter complained of a software issue with a cobas u 411 urine analyzer.The customer alleged the sequence number of the patient¿s sample was changed leading to a mismatch in patient results.The specific results were not provided.The customer reviewed the urine sediment from the test and performed a visual verification where the issue was identified.
 
Additional Manufacturer Narrative
The u 411 analyzer serial number was (b)(6).The investigation is ongoing.
 
Additional Manufacturer Narrative
The field application specialist (fas) confirmed that on the day of the event, there was an issue with the result sequence and the result transmission.No data or actual results were provided for the investigation.Based on other information provided, the investigation noted there were some instrument issues when placing the test strip onto the test strip tray.The field service engineer (fse) replaced the printed circuit boards (pcb) for the test strip detector and the strip check.After the service visit, the customer had no further issues.As no specific data was provided for the investigation, the cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS U 411 URINE ANALYZER
Common Device Name
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17494769
Report Number1823260-2023-02572
Device Sequence Number15277286
Product Code KQO
UDI-Device Identifier04015630925636
UDI-Public04015630925636
Combination Product (Y/N)Y
Initial Reporter CountryCS
PMA/510(K) Number
K093555
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04906969001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/13/2023
Supplement Date Received by Manufacturer08/25/2023
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date09/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-