| Model Number |
FB-21K-1 |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported to olympus, the biopsy forceps cup could not open or close.The event occurred during preparation for use.The intended diagnostic procedure was completed using a similar device.There was no report of patient or user harm associated with the event.During testing and inspection of the returned device, it was discovered that there was foreign material on the device due to lack of proper reprocessing.This medwatch report is being submitted to capture this reportable malfunction which was identified during the evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation.The customer¿s allegation of the biopsy forceps could not open or close was confirmed.This was due to foreign material adhering to the biopsy forceps.Proper reprocessing was not performed after use, and foreign material remained.Additionally, the following finding was also identified, and is as follows: the cups could not open or close.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely lubricant was not applied properly and some foreign materials remained because the appropriate reprocessing was not performed according to the instruction manual.The root cause of adhesion of foreign materials and insufficient lubrication could not be determined.The cup not opening and closing was caused by a combination of adhesion of foreign materials and insufficient lubrication.The event can be detected/prevented by following the instructions for use which state: - "before use, confirm that the distal tip of the forceps is not corroded, dented or discolored, and that the cups can be opened and closed smoothly.Using forceps that are damaged or not working properly may result in one or more of its components falling off inside the patient.If a component falls off the distal end, or if operation of the cups suddenly becomes more difficult, immediately stop the procedure.Carefully withdraw the forceps together with the endoscope to avoid causing injury within the body cavity.Retrieve any parts that have fallen off inside the patient using another grasping forceps."."reprocess the instrument immediately after use by immersing it in a neutral, low-foaming, medical-grade detergent solution, then following the cleaning and sterilization procedures in this chapter.Failure to reprocess the instrument immediately after use, or using another type of detergent may cause corrosion at the instrument¿s distal end.This could cause components to fall off during use and may interfere with operation of the cups."."lubrication: immerse the insertion portion in the lubricant for 2 to 3 seconds.Remove the instrument from the lubricant.Operate the slider to open and close the cups two or three times.Wipe the exterior of the instrument with a clean, dry lint-free cloth and allow the instrument to air dry.".Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|