BIOSENSE WEBSTER INC NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Catalog Number NI75TCFH
Medical Device Problem Codes Crack (1135); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product investigation was completed.Device evaluation details: the product was returned to biosense webster (bwi) for evaluation.Bwi conducted a visual inspection and deflection evaluation of the returned device.Visual analysis of the returned sample revealed that the peek housing cracked with internal parts exposed.This condition could be related to excessive force or manipulation; however, this cannot be conclusively determined.A deflection test was performed and it was found within specifications, that the catheter was deflecting correctly.No deflection issues were verified during the analysis.The instructions for use contain the following recommendations: inspect all components before use.The event described could not be confirmed as the device performed without any deflection issues; other circumstances or issues that occurred during the use of the device could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number 30950211m, and no internal actions were identified.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number:(b)(4).
 
Event or Problem Description
It was reported that a patient underwent an unspecified ablation procedure with a navi-star¿ thermo-cool¿ electrophysiology catheter and post-procedure the bwi product analysis lab identified that the peek housing was cracked with internal parts exposed.During the procedure, the catheter was unable to deflect or relax completely.A second device was used to complete the operation.There was no adverse event reported on patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17495485
Report Number2029046-2023-01742
Device Sequence Number7821903
Product Code OAD
UDI-Device Identifier10846835000566
UDI-Public10846835000566
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P030031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/13/2025
Device Catalogue NumberNI75TCFH
Device Lot Number30950211M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2023
Initial Report FDA Received Date08/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-