| Catalog Number |
AK-04510-S |
| Medical Device Problem Codes |
Unraveled Material (1664); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medwatch report # mw5119230.
|
| |
|
Event or Problem Description
|
|
Medwatch report received: "unknown age - after achieving arterial flow in the left femoral artery with an 18g needle, the wire was introduced, but when attempting to remove the needle, there was a snagging sensation over the wire, so the wire was then attempted to be removed through the needle without success, again with a sensation of snagging.The wire and needle were then removed together, and upon removing the majority of the wire came out of the puncture site, however the wire appeared to have partially unraveled, and metal threads were still stuck within the vascular puncture site, still attached to the wire.It appears that the wire shredded somehow during this process, though the mechanism is unclear.Wires should be able to be inserted and removed through the needle without damage to the wire.The metal threads were removed from the vascular access site.The main concern was how the metal threads were shredded off the wire.".
|
| |
|
Event or Problem Description
|
|
Medwatch report received: "unknown age - after achieving arterial flow in the left femoral artery with an 18g needle, the wire was introduced, but when attempting to remove the needle, there was a snagging sensation over the wire, so the wire was then attempted to be removed through the needle without success, again with a sensation of snagging.The wire and needle were then removed together, and upon removing the majority of the wire came out of the puncture site, however the wire appeared to have partially unraveled, and metal threads were still stuck within the vascular puncture site, still attached to the wire.It appears that the wire shredded somehow during this process, though the mechanism is unclear.Wires should be able to be inserted and removed through the needle without damage to the wire.The metal threads were removed from the vascular access site.The main concern was how the metal threads were shredded off the wire.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The customer description states that the guide wire was attempted to be retracted back through the needle after advancing.This directly contradicted the product ifu, which explicitly cautions the user against withdrawing the guide wire against the needle bevel while in the vessel.A device history record review was performed, and no relevant findings were identified.Based on these circumstances, use error likely caused or contributed to this event.An in-service has been requested to instruct the customer on proper use of the device.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|