• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH SOMATOM DEFINITION FLASH; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH SOMATOM DEFINITION FLASH; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 10430603
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to siemens that an adverse event occurred while operating the somatom definition flash ct system.During a scan of a 4-year-old female patient, the operator reported artifacts on the head images.As a result, the clinical staff made the decision to conduct a rescan of the patient.No negative health consequences were reported for the child associated with the reported event.Therefore, this report is submitted with an abundance of caution as siemens considers children as a vulnerable patient group.
 
Additional Manufacturer Narrative
Siemens has completed an evaluation of the event.The child was rescanned due to image artifacts (windmill artifacts) during a head scan.Such artifacts point to a mismatch between the table movement speed and the scanner rotation speed.The customer adjusted the pitch parameter and rescanned the patient and artifacts no longer appeared.Based on available information the system worked as specified.No further measures are deemed to be necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM DEFINITION FLASH
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1~or
rittigfeld 1
forchheim 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1~or
rittigfeld 1
forchheim 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
4843231631
MDR Report Key17497603
Report Number3004977335-2023-00084
Device Sequence Number16780162
Product Code JAK
UDI-Device Identifier04056869006963
UDI-Public04056869006963
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K230421
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number10430603
Device Catalogue Number10430603
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Initial Report FDA Received Date08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age4 YR
Patient SexFemale
-
-