Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
|
Adding udi # (b)(4).This is a follow up report to add this information.Returned qty.(b)(4).Insert, 80799, 22277-00084203, de, beyond useful life.Test method / gp005-wi02.Inductance: gauge id# 5349, due: (b)(6) 2024, result: 3.15.Meets spec, does not meet spec, n/a.Tip condition: - meets spec, does not meet spec, n/a.Stack condition , meets spec, does not meet spec, n/a.Tested on: g136 gage id# 9626-2, due date: (b)(4) 2023, set pressure: 35 psi.Water flow: output rate: (b)(4) - meets spec , does not meet spec, n/a.Spray pattern: - meets spec , does not meet spec, n/a.Water leak: hp sealing ring, proximal ring, distal ring, seam leak, n/a.Vibration: - meets spec , does not meet spec, n/a.Grip condition: meets spec, does not meet spec, n/a.Other visual observation: insert is good, no fault found.Also insert was tested on a digital thermometer i.D.#6112, due date: (b)(4) 2023.The temperature was 82.3° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed - not confirmed.
|