| Catalog Number |
153622 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that temperatures varied up and down during measurement while using the extension cord for the temperature sensing catheter.The problem continued to occur even after replacing the cable with a new one.The monitor was nihon kohden.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is unconfirmed.Photo samples were submitted, however, could not be evaluated for the reported failure.The extension cable returned without the original packaging.Gross visual evaluation: no damage to the two cords was noted.Connections of the cords to the plug and socket were secure.Using a multimeter, the cord was tested for and found to have continuity.The cord was connected to an in-house temperature sensing catheter submerged in a water bath set at 37 degrees celsius, the cord was then attached to a kilo device and the temperature displayed 37.0 degrees celsius and then stabilized at 37.1 degrees celsius, similar to the thermometer reading.The samples meet the specification "plug and jack must be firmly secured to wire.".The device history record review could not be performed without a lot number.The reported event is unconfirmed a labeling review is not required.H11: section a through f the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that temperatures varied up and down during measurement while using the extension cord for the temperature sensing catheter.The problem continued to occur even after replacing the cable with a new one.The monitor was nihon kohden.
|
| |
|