ZIMMER BIOMET, INC. HEADED SCREW 48 MM LENGTH; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Fracture (1260); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery, the head broke off of the screw and it became stuck in the screwdriver.The surgeon had to remove the screw from the femoral cutting jig with another instrument.The surgical technique was utilized.No foreign bodies were retained.There was no surgical delay.The surgery was not completed with another device.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: 00598304900 - screw inserter/extractor - unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4), this follow-up report is being submitted to relay additional and corrected information.Visual inspection of the returned item found the hex feature is fractured off in the hex feature of the screw inserter/extractor.The rest of this screw wasn¿t returned.Review of the device history record and supplier dhr identified no related deviations or anomalies during manufacturing.Medical records were not provided.Root cause cannot be determined for the screw fracture.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADED SCREW 48 MM LENGTH
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17498302
Report Number0001822565-2023-02130
Device Sequence Number12360322
Product Code LXH
UDI-Device Identifier00889024218277
UDI-Public(01)00889024218277(17)330130(10)65633829
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00598304048
Device Lot Number65633829
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Supplement Date Received by Manufacturer12/15/2023
Initial Report FDA Received Date08/10/2023
Supplement Report FDA Received Date12/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
..
Patient SexMale
-
-